Image quality assessment of low-dose protocols in cone beam computed tomography of the anterior maxilla

Oral Surg Oral Med Oral Pathol Oral Radiol. 2022 Apr;133(4):483-491. doi: 10.1016/j.oooo.2021.10.001. Epub 2021 Oct 11.

Abstract

Objectives: To evaluate overall image quality and visibility of anatomic structures on low-dose cone beam computed tomography (CBCT) scans and the effect of a noise reduction filter for assessment of the anterior maxilla.

Methods: We obtained 48 CBCT volumes on 8 skull-phantoms using 6 protocols: 2 clinical default protocols [standard definition (SD) and high definition (HD)] and 4 low-dose protocols, 2 with a noise reduction filter [ultra-low-dose with high definition (ULDHD) and ultra-low-dose (ULD)] and 2 without [low-dose with high definition (LDHD) and low-dose (LD)]. Overall image quality and visibility of 8 anatomic structures were assessed by 5 observers and statistically analyzed using the Wilcoxon signed rank test. Intra- and interobserver agreement was measured using Cohen's weighted kappa.

Results: HD provided higher overall image quality than diagnostically required; LD scored lower than diagnostically acceptable. ULDHD, ULD, and LDHD were acceptable. For anatomic structures, ULDHD and ULD were acceptable. LDHD and LD showed significantly inferior visibility for 1 and 4 structures, respectively. Mean values of intra- and interobserver agreement were 0.395 to 0.547 and 0.350 to 0.370, respectively.

Conclusions: ULDHD, ULD, and LDHD may be recommended for assessment of impacted maxillary canines. The noise reduction filter affects image quality positively only at low exposure.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cone-Beam Computed Tomography* / methods
  • Humans
  • Maxilla* / diagnostic imaging
  • Phantoms, Imaging
  • Radiation Dosage
  • Skull