The established and the challenger: A direct comparison of current cryoballoon technologies for pulmonary vein isolation

J Cardiovasc Electrophysiol. 2022 Jan;33(1):48-54. doi: 10.1111/jce.15288. Epub 2021 Nov 18.

Abstract

Introduction: Cryoballoon (CB) ablation for pulmonary vein isolation (PVI) is an effective treatment of atrial fibrillation (AF). Recently, a novel cryoablation system was introduced. The aim of the study was to compare the safety, efficacy and biophysical characteristics of a novel cryoablation system (POLARx™; Boston Scientific) to a commonly used and clinically well characterized system (Arctic Front Advance Pro™, AFA; Medtronic).

Methods and results: Fifty consecutive patients with symptomatic AF, who underwent CB-based ablation with the POLARx were compared to 50 consecutive patients treated with the AFA. Acute PVI was achieved in 99.8% (POLARx 99.5%, AFA 100%, p = 1.00). Time to isolation (TTI) was comparable in both groups (POLARx 35 [27, 48] s, AFA 30 [21, 43] s, p = 0.165). The POLARx showed a lower balloon temperature at TTI (POLARx -44 [-50, -36] °C, AFA -31 [-38, -21] °C, p < 0.001) and lower nadir temperature (POLARx -60 [-65, -55] °C, AFA -48 [-54, -45] °C, p < 0.001). Procedure time (POLARx 80 [60, 105] min, AFA 62 [42, 80] min, p < 0.001), fluoroscopy time (POLARx 17 [13, 22] min, AFA 11 [7, 16] min, p < 0.001) and freeze cycles per patient (POLARx 5 [4, 6], AFA 4.5 [4, 5], p = 0.002) were higher in the POLARx group. Two cerebral ischemic events occurred in the POLARx group, two patients in each group had phrenic nerve injury.

Conclusion: Both systems enable effective isolation of pulmonary veins. The POLARx required longer procedure and fluoroscopy times. Larger, prospective and randomized studies are needed to assess long-term efficacy and safety of this technology.

Keywords: atrial fibrillation; catheter ablation; cryoballoon technology; pulmonary vein isolation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Atrial Fibrillation* / diagnostic imaging
  • Atrial Fibrillation* / surgery
  • Catheter Ablation* / adverse effects
  • Cryosurgery* / adverse effects
  • Cryosurgery* / methods
  • Humans
  • Prospective Studies
  • Pulmonary Veins* / diagnostic imaging
  • Pulmonary Veins* / surgery
  • Technology