Objective: To assess the prevalence and implications of phase changes in extracochlear electrocochleography (ECochG) recordings during cochlear implantation.
Materials and methods: Extracochlear ECochG recordings were performed before and after insertion of the cochlear implant (CI) electrode by a recording electrode placed on the promontory. Acoustic stimuli were tone bursts at 250, 500, 750, and 1,000 Hz. The pure tone average (PTA) was determined before and approximately 4 weeks after surgery.
Results: Extracochlear ECochG recordings in 69 ears of 68 subjects were included. At 250 Hz, the mean phase change was 43° (n = 50, standard deviation (SD) 44°), at 500 Hz 36° (n = 64, SD 36°), at 750 Hz 33° (n = 42, SD 39°), and at 1,000 Hz 22° (n = 54, SD 27°). Overall, in 48 out of 210 ECochG recordings a phase change of ≥45° (23%) was detectable. Ears with an amplitude drop >3 dB and a phase change ≥45° (n = 3) had a complete or near complete loss of residual cochlear function in all cases. A phase change of ≥90° in one recording was not associated with a larger amplitude change of the ECochG signal (1.9 dB vs. -0.9 dB, p = 0.1052, n = 69), but with a significantly larger postoperative hearing loss (17 dB vs. 26 dB, p = 0.0156, n = 69).
Conclusions: Phase changes occur regularly in extracochlear ECochG recordings during cochlear implantation. Phase changes of ≥90° with or without amplitude changes in the ECochG signal are associated with a larger postoperative hearing loss and could therefore represent an independent marker for cochlear trauma or changes of inner ear mechanics relevant for the postoperative hearing outcome.