Closed-Loop Insulin Delivery Versus Sensor-Augmented Pump Therapy in Older Adults With Type 1 Diabetes (ORACL): A Randomized, Crossover Trial

Diabetes Care. 2022 Feb 1;45(2):381-390. doi: 10.2337/dc21-1667.

Abstract

Objective: To assess the efficacy and safety of closed-loop insulin delivery compared with sensor-augmented pump therapy among older adults with type 1 diabetes.

Research design and methods: This open-label, randomized (1:1), crossover trial compared 4 months of closed-loop versus sensor-augmented pump therapy. Eligible adults were aged ≥60 years, with type 1 diabetes (duration ≥10 years), using an insulin pump. The primary outcome was continuous glucose monitoring (CGM) time in range (TIR; 3.9-10.0 mmol/L).

Results: There were 30 participants (mean age 67 [SD 5] years), with median type 1 diabetes duration of 38 years (interquartile range [IQR] 20-47), randomized (n = 15 to each sequence); all completed the trial. The mean TIR was 75.2% (SD 6.3) during the closed-loop stage and 69.0% (9.1) during the sensor-augmented pump stage (difference of 6.2 percentage points [95% CI 4.4 to 8.0]; P < 0.0001). All prespecified CGM metrics favored closed loop over the sensor-augmented pump; benefits were greatest overnight. Closed loop reduced CGM time <3.9 mmol/L during 24 h/day by 0.5 percentage points (95% CI 0.3 to 1.1; P = 0.0005) and overnight by 0.8 percentage points (0.4 to 1.1; P < 0.0001) compared with sensor-augmented pump. There was no significant difference in HbA1c between closed-loop versus sensor-augmented pump stages (7.3% [IQR, 7.1-7.5] (56 mmol/mol [54-59]) vs. 7.5% [7.1-7.9] (59 mmol/mol [54-62]), respectively; P = 0.13). Three severe hypoglycemia events occurred during the closed-loop stage and two occurred during the sensor-augmented pump stage; no hypoglycemic events required hospitalization. One episode of diabetic ketoacidosis occurred during the sensor-augmented pump stage; no serious adverse events occurred during the closed-loop stage.

Conclusions: Closed-loop therapy is an effective treatment option for older adults with long-duration type 1 diabetes, and no safety issues were identified. These older adults had higher TIR accompanied by less time below range during closed loop than during sensor-augmented pump therapy. Of particular clinical importance, closed loop reduced the time spent in hypoglycemic range overnight.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Blood Glucose Self-Monitoring
  • Blood Glucose*
  • Cross-Over Studies
  • Diabetes Mellitus, Type 1* / chemically induced
  • Diabetes Mellitus, Type 1* / drug therapy
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin
  • Insulin Infusion Systems
  • Middle Aged

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin

Associated data

  • ANZCTR/ACTRN12619000515190
  • figshare/10.2337/figshare.16924480