Intraoperative sensing increase predicts long-term pacing threshold in leadless pacemakers

J Interv Card Electrophysiol. 2022 Apr;63(3):679-686. doi: 10.1007/s10840-021-01111-x. Epub 2022 Jan 4.

Abstract

Purpose: High pacing threshold (HPT) and very high pacing threshold (VHPT) are known to have a negative impact on leadless pacemaker battery longevity, representing the most common reason for device repositioning. In this study, we evaluated if intraoperative electrical parameters recorded during Micra™ VR implant would be able to predict device performance during follow-up (FU).

Methods: A total of 93 patients undergoing Micra™ VR implant were retrospectively considered. Patients were enrolled in the study if electrical assessment was performed at least twice at implant, at Micra™ final positioning and after removal of the delivery system. All patients received a FU visit at 1 and 12 month after discharge. R-wave sensing amplitude, pacing threshold (PT), and impedance were recorded at each visit.

Results: When compared to the first assessment, R-wave sensing amplitude increased by 19.1% after 13 ± 4 min (+ 1.71 ± 0.2 mV, 95% CI 1.4 to 2.02; p < .001). Conversely, there was a significant PT decrease of 22.1% at 12-month FU (- 0.22 ± 0.03 V, 95% CI - 0.13 to - 0.31; p < .001). Among patients with HPT, acute increase of R-wave sensing of 1.5 mV after 14 ± 4 min predicted a significant reduction of PT below 1 V/0.24, at 12-month post-implant (R = 0.72, 95% CI 0.13 to 0.33, p < .001), with a sensitivity of 87.5% (95% CI 0.61-0.98) and a specificity of 88.8% (95% CI 0.51-0.99).

Conclusion: A 1.5-mV increase in R-wave amplitude at implant is predictive of PT normalization (< 1.0 V/0.24 ms) at 12-month FU. This finding may have practical implications for device repositioning in case of HPT recording at implant.

Keywords: High pacing threshold; Leadless pacemaker; Micra-VR intraoperative sensing increasing; Predictor of threshold improvement.

MeSH terms

  • Cardiac Pacing, Artificial*
  • Equipment Design
  • Humans
  • Pacemaker, Artificial*
  • Retrospective Studies
  • Treatment Outcome