The PRONTO study: Clinical performance of ID NOW in individuals with compatible SARS-CoV-2 symptoms in walk-in centres-accelerated turnaround time for contact tracing

Can Commun Dis Rep. 2021 Dec 9;47(12):534-542. doi: 10.14745/ccdr.v47i12a04.

Abstract

Background: This PRONTO study investigated the clinical performance of the Abbott ID NOWTM (IDN) COVID-19 diagnostic assay used at point of care and its impact on turnaround time for divulgation of test results.

Methods: Prospective study conducted from December 2020 to February 2021 in acute symptomatic participants presenting in three walk-in centres in the province of Québec.

Results: Valid paired samples were obtained from 2,372 participants. A positive result on either the IDN or the standard-of-care nucleic acid amplification test (SOC-NAAT) was obtained in 423 participants (prevalence of 17.8%). Overall sensitivity of IDN and SOC-NAAT were 96.4% (95% CI: 94.2-98.0%) and 99.1% (95% CI: 97.6-99.8), respectively; negative predictive values were 99.2% (95% CI: 98.7-99.6%) and 99.8% (95% CI: 99.5-100%), respectively. Turnaround time for positive results was significantly faster on IDN.

Conclusion: In our experience, IDN use in symptomatic individuals in walk-in centres is a reliable sensitive alternative to SOC-NAAT without the need for subsequent confirmation of negative results. Such deployment can accelerate contact tracing, reduce the burden on laboratories and increase access to testing.

Keywords: Abbott ID NOW; COVID-19; Canada; SARS-CoV-2; diagnostic performance; nucleic acid amplification tests; point-of-care testing; predictive value; rapid tests; sensitivity and specificity.