Introduction: The Organ Care System (OCS) is an ex vivo perfusion platform for donor heart preservation. Short/mid-term post-transplant outcomes after its use are comparable to standard cold storage (CS). We evaluated long-term outcomes following its use.
Methods: Between 2011 and 2013, 38 patients from a single center were randomized as a part of the PROCEED II trial to receive allografts preserved with CS (n = 19) or OCS (n = 19). Endpoints included 8-year survival, survival free from graft-related deaths, freedom from cardiac allograft vasculopathy (CAV), non-fatal major adverse cardiac events (NF-MACE), and rejections.
Results: Eight-year survival was 57.9% in the OCS group and 73.7% in the CS group (p = .24). Freedom from CAV was 89.5% in the OCS group and 67.8% in the CS group (p = .13). Freedom from NF-MACE was 89.5% in the OCS group and 67.5% in the CS group (p = .14). Eight-year survival free from graft-related death was equivalent between the two groups (84.2% vs. 84.2%, p = .93). No differences in rejection episodes were observed (all p > .5).
Conclusions: In select patients receiving OCS preserved allografts, late post-transplant survival trended lower than those transplanted with an allograft preserved with CS. This is based on a small single-center series, and larger numbers are needed to confirm these findings.
Keywords: heart (allograft) function/dysfunction; heart disease; organ perfusion and preservation; patient survival.
© 2022 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.