Purpose: To investigate the diagnostic contribution of grating visual acuity (GVA) measured by sweep pattern-reversal visually evoked potentials (SPRVEP) in unexplained visual loss (UVL).
Methods: This case-control study included adult patients under suspicion of UVL referred to SPRVEP and transient pattern-reversal visually evoked potentials (TPRVEP) testing. Optotype visual acuity (OVA) was measured by ETDRS 4-meter chart and GVA by SPRVEP. UVL patients were assigned into three distinctive categories, according to the presence of ocular disease, motivation, and electrophysiological evaluation, as follows: exaggerators, malingerers, and psychogenic. Healthy controls and patients with organic visual loss were also tested. Receiver operating characteristic (ROC) curve was constructed to evaluate the diagnostic performance of GVA and TPRVEP parameters.
Results: A total of 76 patients with UVL were analyzed: 60 (79.0%) exaggerators, 11 (14.4%) malingerers, and 5 (6.6%) psychogenic. Controls were 49 subjects evaluated for TPRVEP and 28 subjects for SPRVEP. There were 13 patients with organic visual loss enrolled. Mean difference between OVA and GVA was 1.19±0.67 (median=0.84; 95% CI: 1.04 to 1.34) in UVL and 0.14 ±0.09 (median= 0.14; 95% CI: 0.08 to 0.20) in organic visual loss. The area under the ROC curve (AUC) of GVA to distinguish UVL from healthy controls was 0.998 with a cutoff of 0.09 logMAR showing specificity of 100% and sensitivity of 96.0%.
Conclusions: GVA measured by SPRVEP had good diagnostic validity to discriminate patients with unexplained visual loss from healthy controls and patients with organic visual loss, demonstrating its contribution to the diagnosis of this condition.
Keywords: Adults; Malingering; Sweep VEP; Unexplained visual loss; Visual acuity; Visual evoked potentials.
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.