Bladder filling in patients undergoing prostate radiotherapy on a MR-linac: The dosimetric impact

Tech Innov Patient Support Radiat Oncol. 2022 Feb 16:21:41-45. doi: 10.1016/j.tipsro.2022.02.002. eCollection 2022 Mar.

Abstract

The implementation of adaptive radiotherapy for prostate cancer compensates for inter-fraction motion, at the penalty of increased time in room. The subsequent increase in bladder filling may impact dosimetry, which we have investigated on ten patients treated on the MR-linac. Patients drank water before treatment, to achieve a bladder volume of 200-300 cm3. Bladder and bowel were re-outlined offline on 140 images and plans recalculated. All mandatory bladder dose constraints and 99.1% of bowel constraints were achieved at time of treatment, despite varying bladder volumes and varying adherence to original bladder filling guidance.

Keywords: Bladder filling; MR-linac; Prostate; Radiotherapy.