Therapeutic drug monitoring of carbamazepine is necessary for its clinical application with the purpose to reach therapeutic concentration and reduce the risk of concentration-dependent toxicity. An ultrafast analytical assay for quantification of carbamazepine in human plasma was developed and validated based on direct analysis in real time tandem mass spectrometry (DART-MS/MS). After reversed phase solid phase extraction with Waters Oasis HLB, carbamazepine and internal standard carbamazepine-D2N15 were monitored by positive ion mode followed by multiple reaction monitoring (MRM) of the transitions at m/z 237.1→194.0 and 240.1→196.2, respectively. The DART-MS/MS method is ultrafast and high-throughput and the analytical time for each sample is only 0.4 min. The assay was linear in the concentration range 0.50-30 μg/mL and intra- and inter-day accuracies were within ± 15% and trueness were < 13.9% at all concentrations. The method was successfully applied to therapeutic drug monitoring of carbamazepine in human plasma.
Keywords: Carbamazepine; DART-MS/MS; Therapeutic drug monitoring.