152Eu has been standardized by three independent 4π β-γ coincidence counting systems with beta detectors as proportional counter, plastic scintillator and liquid scintillator along with the CIEMAT/NIST method. The average activity concentration by primary methods was linked to key comparison reference value (KCRV) by comparing it with that of 4π γ ionization chamber (GIC) whose calibration factor was determined from the KCRV (BIPM.RI(II)-K1.Eu-152 and CCRI(II)-K2.Eu-152) and deviates from GIC by ± 0.16% indicating good agreement within standard uncertainties.
Keywords: 152Eu; 4π β-γ coincidence technique; CCRI(II)–K2.Eu-152; CIEMAT/NIST method; Ionization chamber; Plastic scintillator; Standardization.