Clinical outcomes in a U.S. registration study of a new EDOF intraocular lens with a nondiffractive design

J Cataract Refract Surg. 2022 Nov 1;48(11):1297-1304. doi: 10.1097/j.jcrs.0000000000000978. Epub 2022 May 26.

Abstract

Purpose: To evaluate the effectiveness and safety of the DFT015 intraocular lens (IOL) (AcrySof IQ Vivity Extended Vision) compared with an aspheric monofocal control IOL (AcrySof IQ model SN60WF).

Setting: 11 investigation sites in the U.S.

Design: Prospective randomized controlled clinical study.

Methods: Patients aged 22 years or older with bilateral cataracts were randomized to receive bilateral implantation of DFT015 or SN60WF. The 4 coprimary effectiveness outcomes (6 months postoperatively) were monocular photopic distance-corrected intermediate visual acuity (DCIVA), monocular photopic corrected distance visual acuity (CDVA), monocular depth of focus (DoF), and the percentage of patients achieving a DCIVA of 0.2 logMAR or better. The mean monocular photopic distance-corrected near visual acuity (DCNVA) was a secondary effectiveness outcome. Safety and patient-reported visual disturbances were evaluated through questionnaires.

Results: 218 patients (435 eyes) completed the study. Compared with SN60WF, DFT015 demonstrated superior mean monocular photopic DCIVA ( P < .001), noninferior mean monocular photopic CDVA, and superior mean monocular photopic DCNVA ( P < .001) and provided an extended monocular DoF (increase of 0.54 diopters at 0.2 logMAR). With DFT015, 78 first eyes (72.9%) achieved a DCIVA of 0.2 logMAR or better at 6 months. Incidences of ocular serious adverse events and patient-reported most bothersome visual disturbances were low and consistent between groups.

Conclusions: DFT015 is safe and effective for the visual correction of aphakia, exceeding American National Standards Institute criteria for an extended depth-of-focus IOL by providing superior DCIVA and DCNVA, with comparable CDVA and visual disturbances to the SN60WF monofocal IOL.

Trial registration: ClinicalTrials.gov NCT03274986.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Phacoemulsification*
  • Prospective Studies
  • Prosthesis Design
  • Pseudophakia
  • Refraction, Ocular

Associated data