Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results

Diagn Microbiol Infect Dis. 2022 Aug;103(4):115726. doi: 10.1016/j.diagmicrobio.2022.115726. Epub 2022 May 20.

Abstract

This study used digital polymerase chain reaction (dPCR) to determine whether envelope (E) gene-negative and nucleocapsid (N2) gene-positive (E-N+) results obtained with the Cepheid Xpert Xpress SARS-CoV-2 assay are reliable. Using droplet digital PCR results as a reference, 18 of 22 E-N+ samples with a low viral load (81.8%) were identified as true positives.

Keywords: Cepheid Xpert Xpress SARS-CoV-2 assay; E gene negative; N2 gene positive; SARS-CoV-2; digital PCR; low viral load.

MeSH terms

  • COVID-19 Testing
  • COVID-19* / diagnosis
  • Clinical Laboratory Techniques / methods
  • Humans
  • Nasopharynx
  • Nucleocapsid / genetics
  • Polymerase Chain Reaction
  • Reproducibility of Results
  • SARS-CoV-2* / genetics
  • Sensitivity and Specificity