Background: The IL-3-pSTAT5 assay, a new, rapid, and standardized flow-cytometry-based assay may compensate for several limitations of the colony-forming unit (CFU) assay typically used for stem cell potency assessments of cord blood units (CBU). We performed an inter-laboratory evaluation of the performance of this new assay.
Study design and methods: This Biomedical Excellence for Safer Transfusion (BEST) Collaborative multicenter, international study included 15 participants from public cord blood banks (CBBs), CBB-supporting research laboratories, and stem cell laboratories. To perform the IL-3-pSTAT5 assay, participating centers received reagents, instructions, and 10 blind CBU samples, including eight normal samples and two samples exposed to a transient warming event. We measured inter-laboratory agreement qualitatively (proportion of correctly classified samples) and quantitatively (coefficient of variation [CV], correlation coefficients, receiver operating characteristics (ROC) curve, and intraclass correlation coefficient [ICC]).
Results: The qualitative agreement was 97.3% (i.e., 107/110; Fleiss' kappa = 0.835). The average CV on a per-sample basis was 11.57% among all samples, 8.99% among normal samples, and on a per-center basis was 9.42% among normal samples. In a correlation matrix that compared results across centers, the mean Pearson's correlation coefficient was 0.88 (standard deviation = 0.04). The ICC was 0.83 (95% confidence interval = 0.68-0.95). The area under the curve (AUC) from the ROC curve was 0.9974.
Discussion: Excellent qualitative and quantitative agreement was exhibited across laboratories. The IL-3-pSTAT5 assay may therefore be implemented in flow cytometry laboratories to rapidly and reliably provide standardized measures of stem cell potency in CBUs.
Keywords: cord blood; flow cytometry; multicenter evaluation; potency; stem cells.