Background: The new noninvasive finger sensor system NICCI (Getinge; Gothenburg, Sweden) allows continuous cardiac output monitoring. We aimed to investigate its cardiac output measurement performance.
Objectives: To investigate the NICCI system's cardiac output measurement performance.
Design: Prospective method comparison study.
Setting: University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Patients: Fifty-one patients after cardiac surgery.
Main outcome measures: We performed a method comparison study in 51 patients after cardiac surgery to compare NICCI cardiac output (CO NICCI ) and NICCI cardiac output calibrated to pulmonary artery thermodilution cardiac output measurement (CO NICCI-CAL ) with pulmonary artery thermodilution cardiac output (CO PAT ). As a secondary analysis we also compared CNAP cardiac output (CO CNAP ) and externally calibrated CNAP cardiac output (CO CNAP-CAL ) with CO PAT .
Results: We analysed 299 cardiac output measurement pairs. The mean of the differences (95% limits of agreement) between CO NICCI and CO PAT was 0.6 (-1.8 to 3.1) l min -1 with a percentage error of 48%. The mean of the differences between CO NICCI-CAL and CO PAT was -0.4 (-1.9 to 1.1) l min -1 with a percentage error of 29%. The mean of the differences between CO CNAP and CO PAT was 1.0 (-1.8 to 3.8) l min -1 with a percentage error of 53%. The mean of the differences between CO CNAP-CAL and CO PAT was -0.2 (-2.0 to 1.6) l min -1 with a percentage error of 35%.
Conclusion: The agreement between CO NICCI and CO PAT is not clinically acceptable.
Trial registration: The study was registered in the German Clinical Trial Register (DRKS00023189) after inclusion of the first patient on October 2, 2020.