Transcatheter vacuum-assisted left-sided mass extraction with the AngioVac system

Catheter Cardiovasc Interv. 2022 Oct;100(4):628-635. doi: 10.1002/ccd.30345. Epub 2022 Jul 28.

Abstract

Objectives: To study the safety and efficacy of AngioVac for left-sided transcatheter vacuum-assisted mass extraction (TVME).

Background: The AngioVac system is approved for right-sided TVME and has emerged as an effective and safe alternative for open surgical treatment. The use of the AngioVac device for aspiration of left-sided TVME has been limited.

Methods: Consecutive patients from two Michigan centers who underwent left-sided TVME were included. Data on patient demographics, procedural information, in-hospital and follow-up events were collected through electronic medical records review. Technical success was defined as aspirating of 70%-100% of the material.

Results: Ten patients (mean age 58.3 [±17.3] years, 50% male) were included. Indications for TMVE were in large for recurrent embolic events. All patients underwent bilateral cerebro-embolic protection using the Sentinel device. The total mean procedure time was 192.5 (±47.5) min of which the meantime for active aspiration (bypass time) was 9.3 (±4.2) min. The circuit configuration was: arteriovenous (AV) in four cases and arterioarterial (AA) in six cases. Successful aspiration was achieved in 80% of cases. No complications were reported (range follow-up 1-16 months).

Conclusions: Our small case series demonstrates the feasibility and safety of the AngioVac system in left-sided mass extraction. Larger trials are needed to further demonstrate its effectiveness and safety and potentially apply for on-label use.

Keywords: CATT-catheterization; ENDO-endocarditis; THRO-thrombectomy; transseptal.