The jugular approach for leadless pacing: A novel and safe alternative

Pacing Clin Electrophysiol. 2022 Oct;45(10):1248-1254. doi: 10.1111/pace.14587. Epub 2022 Sep 12.

Abstract

Aims: To evaluate safety of leadless pacemaker implantation through the internal jugular vein in a larger cohort with longer follow-up. Moreover, feasibility of non-apical pacing as well as relation between pacing site and QRS duration were assessed.

Methods: Eighty Two consecutive patients, who received a leadless pacemaker though the internal jugular vein, were included. Electrical parameters were measured at regular follow-up and any complications were registered. Paced QRS interval was compared for three pacing sites, RVOT, RV mid septum, and RV apical septum.

Results: In all patients, the leadless pacemaker was implanted successfully. In 69 patients, the device was implanted in a non-apical position. In 71% of cases, the device could be deployed at first attempt. The median fluoroscopy time was 4.4 min (range 0.9-51) The paced QRS interval was significantly narrower for non-apical pacing sites compared to apical pacing si 156 vs. 179 ms. p = .04, respectively. During mean follow-up of 16 months (range 0-43 months), electrical parameters remained stable. Two complications occurred, which could be resolved during the implant procedure. There were no access site related complications.

Conclusion: The jugular approach for leadless pacemaker implantation is feasible and may avoid vascular complications. It facilitates non-apical positioning of leadless pacemakers leading to a narrower paced QRS interval. The jugular approach allows for immediate post procedural ambulation.

Keywords: AV-block; MICRA pacemaker; bradycardia; internal jugular vein; leadless pacing; transcatheter pacemaker.

MeSH terms

  • Cardiac Pacing, Artificial
  • Equipment Design
  • Humans
  • Jugular Veins
  • Pacemaker, Artificial*
  • Treatment Outcome