Objectives: We have proposed a novel intramedullary nail (Ni-Nail) by incorporating a sustentaculum tali screw to improve the fixation stability of minimally invasive treatment for calcaneal fractures. This study aimed to evaluate the biomechanical characters of the Ni-Nail system and compare it with traditional C-Nail system.
Methods: A finite element model of a Sanders type-IIIAB calcaneal fracture was reconstructed and fixed using two intramedullary nail systems, which was validated by a cadaver study. A vertical loading of 700 N was applied to the subtalar joint surfaces, and 525 N Achilles tendon tension was applied to the superior border of the Achilles tuberosity. The von Mises stresses and fracture displacements of both fixation models were evaluated.
Results: The maximum von Mises stress of the screws of Ni-Nail and C-Nail were 27.92 MPa and 57.42 MPa, respectively, while that of the main nail were 67.44 MPa and 53.01 MPa. In addition, the maximum fracture displacement of the Ni-Nail was larger than that of C-Nail by 15.6% (0.37 mm vs.0.32 mm).
Conclusions: Our static simulation analysis showed that both Ni-Nail and C-Nail demonstrated similar biomechanical stability for calcaneal fixation. The Ni-Nail features a simple structure that is easier to operate and less traumatizing. Future studies may consider to further evaluate the clinical effectiveness by clinical trials and follow-ups.
Keywords: Biomechanics; Calcaneal fracture; Finite element analysis; Fracture fixation; Intramedullary nail.
Copyright © 2022. Published by Elsevier Ltd.