Purpose: To evaluate the safety, primary patency, and clinical outcomes of hepatic artery stent graft (SG) placement for vascular injuries.
Materials and methods: Patients treated with hepatic arterial SG placement for vascular injuries between September 2018 and September 2021 were reviewed. Data on demographic characteristics, indication, stent graft characteristics, antiplatelet/anticoagulant use, clinical success rate, complications, and type of follow-up imaging were collected. Follow-up images were reviewed by 2 independent reviewers to assess primary patency. A time-to-event analysis was performed. The median duration of stent graft patency was estimated using Kaplan-Meier curves. A Cox proportional hazard model was used to evaluate factors related to stent graft patency.
Results: Thirty-five patients were treated with hepatic arterial SG placement, 11 for postoperative bleeds and 24 for hepatic artery infusion pump catheter-related complications. Clinical success was achieved in 32 (91%) patients (95% CI, 77-98). The median primary patency was 87 days (95% CI, 73-293). Stent grafts of ≥6-mm diameter retained patency for a longer duration than that with stent grafts of smaller diameters (6 mm vs 5 mm; hazard ratio, 0.35; 95% CI, 0.14-0.88; P = .026; and 7+ mm vs 5 mm; hazard ratio, 0.27; 95% CI, 0.09-0.83; P = .023). Anticoagulation/antiplatelet regimen was not associated with increased stent graft patency duration (P > .05). Only minor complications were reported in 2 (5.7%) patients.
Conclusions: Stent grafts can be used safely and effectively to treat injuries of the hepatic artery. Stent graft diameters of ≥6 mm seem to provide more durable patency.
Copyright © 2022 SIR. Published by Elsevier Inc. All rights reserved.