"Bowing Technique" for Using a HeartMate 3 Mini-Apical Cuff

ASAIO J. 2023 Apr 1;69(4):e149-e151. doi: 10.1097/MAT.0000000000001830. Epub 2022 Mar 11.

Abstract

A HeartMate 3 mini-apical cuff is designed to facilitate minimally invasive left ventricular assist implantation. A possible problem of using a mini-apical cuff is that multiple buttress braided sutures can interfere with a HeartMate 3 pump when locking the pump. If this happens in a minimally invasive approach, multiple manipulations to engage the pump can cause catastrophic myocardial injury and uncontrollable bleeding, especially in patients with a fragile myocardium. We are reporting a simple and effective technique to prevent suture jamming in the setting of mini-apical cuff implantation utilizing multiple buttress sutures.

MeSH terms

  • Heart-Assist Devices*
  • Hemorrhage*
  • Humans
  • Myocardium