High sensitivity SARS-CoV-2 detection using graphene oxide-multiplex qPCR

Anal Chim Acta. 2022 Nov 22:1234:340533. doi: 10.1016/j.aca.2022.340533. Epub 2022 Oct 18.

Abstract

The emerging pandemic of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) critically challenges early and accurate virus diagnoses. However, the current gold standard for SARS-CoV-2 detection, reverse transcription-quantitative polymerase chain reaction (RT-qPCR), has reportedly failed to detect low-viral loads. One compound, graphene oxide (GO), which adsorbs single-stranded DNA (ssDNA), has been widely applied in molecular pathogen detection. This study presents a highly sensitive GO-multiplex qPCR method for simultaneous detection of two SARS-CoV-2 genes (RdRP and E) and one reference gene (RNase P). In a GO-multiplex qPCR system, GO pre-absorbs each forward primer to form specific GO-forward primer composites before entering the amplification system. Target gene amplification is confined within the primer-enriched composites, thus, improving the sensitivity of the assay. Compared to conventional multiplex qPCR, GO-multiplex qPCR reduces the limit of detection by 10-fold to 10 copies/reaction. Hence, the GO-multiplex qPCR assay can be effectively used for SARS-CoV-2 detection.

Keywords: Detection; Graphene oxide; High sensitivity; Multiplex qPCR; SARS-CoV-2.

MeSH terms

  • COVID-19 Testing
  • COVID-19* / diagnosis
  • Clinical Laboratory Techniques / methods
  • Humans
  • RNA, Viral / analysis
  • RNA, Viral / genetics
  • SARS-CoV-2* / genetics
  • Sensitivity and Specificity

Substances

  • graphene oxide
  • RNA, Viral