Comparison of two radiofrequency-based hemostatic devices: saline-linked bipolar vs. cooled-electrode monopolar

Int J Hyperthermia. 2022;39(1):1397-1407. doi: 10.1080/02656736.2022.2140840.

Abstract

Purpose: To characterize the coagulation zones created by two radiofrequency (RF)-based hemostatic devices: one comprised an internally cooled monopolar electrode and the other comprised externally irrigated bipolar electrodes (saline-linked).

Materials and methods: RF-induced coagulation zones were created on ex vivo and in vivo porcine models. Computer modeling was used to determine the RF power distribution in the saline-linked device.

Results: Both external (irrigation) and internal cooling effectively prevented tissue sticking. Under ex vivo conditions in 'painting' application mode, coagulation depth increased with the applied power: 2.8 - 5.6 mm with the 3-mm monopolar electrode, 1.6 - 6.0 mm with the 5-mm monopolar electrode and 0.6 - 3.2 mm with the saline-linked bipolar electrodes. Under in vivo conditions and using spot applications, the 3-mm monopolar electrode created coagulation zones of similar depth to the saline-linked bipolar electrodes (around 3 mm), while the 5-mm monopolar electrode created deeper coagulations (4.5 - 6 mm) with less incidence of popping. The presence of saline around the saline-linked bipolar electrodes meant that a significant percentage of RF power (50 - 80%) was dissipated by heating in the saline layer. Coagulation zones were histologically similar for all the tested devices.

Conclusions: Both external (irrigation) and internal cooling in hemostatic RF devices effectively prevent tissue sticking and create similar coagulation zones from a histological point of view. Overall, saline-linked bipolar electrodes tend to create shallower coagulations than those created with an internally cooled monopolar electrode.

Keywords: Cooled electrode; hemostatic device; radiofrequency; saline-linked; thermal ablation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Catheter Ablation*
  • Electrodes
  • Equipment Design
  • Hemostatics*
  • Liver / surgery
  • Radio Waves
  • Saline Solution / therapeutic use
  • Swine

Substances

  • Hemostatics
  • Saline Solution