Purpose: To implement a single set-up monthly QA procedure for 9 different beam parameters at different gantry angles and evaluate its clinical implementation over a 12 month period.
Methods: We developed a QA procedure using an array detector (PTW Octavius 1500XDR) embedded in a rotational unit (PTW Octavius 4D) at our proton facility. With a single set-up we can monitor field central axis position, field symmetry, field size, flatness, penumbrae, output, spot size, spot position and range at different gantry angles (AAPM TG 224). The set-up is irradiated with homogenous 2D fields with dynamic aperture and spot patterns at five gantry angles. A modular top is used to check the range consistency. Absolute γ analysis were performed to compare measured dose distributions to calculated dose. All other parameters are directly extracted from the measurements. Additionally, the sensitivity of the set-up to small changes in beam parameters were compared to the Lynx detector (IBA).
Results: Over a 12 month period, output, symmetry, and flatness were within ± 2 %; FWHM, spot positions, penumbra widths, and central axis fields were within ± 1 mm. Range differences were all within 1/2 of the energy spacing (±0.6 MeV) relative to baseline. Most (2 %, 2 mm) γ-analysis showed agreement scores higher than 90 %. The sensitivity is comparable to the Lynx detector and measurement time is reduced by 40 %.
Conclusion: The time-efficient monthly QA procedure that we developed can accurately be used to measure a large range of beam parameters at different gantry angles, within the TG 224 AAPM recommendations.
Keywords: Dosimetry; Ionisation chamber array; Proton therapy; Qa; Rotational phantom.
Copyright © 2022 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.