Liver fatty acid binding protein (L-FABP) is an intercellular lipid chaperone protein that selectively combines with unsaturated free fatty acids and transports them to mitochondria or peroxisomes. L-FABP is a promising biomarker for the early detection of renal diseases in humans. Herein a chemiluminescence method (CLIA) was demonstrated to measure the level of urinary L-FABP in the urinary samples. An anti-(L-FABP)-magnetic beads complex was prepared to capture the analyte target. Sensitivity, precision, accuracy, interference effect, high-dose hook effect of the developed assay were evaluated. Under the suitable experimental parameters, the established method have a wide linear range (0.01-10 ng/mL) and also showed a sufficiently low limit of detection of 0.0060 ng/mL. Besides, the satisfactory recoveries of the method in the urinary were ranged from 97.74%-112.32%, which was well within the requirement of clinical analysis. Furthermore, this proposed method has been successfully applied to the clinical determination of L-FABP in patients who have been diagnosed with kidney disease. The results showed that CLIA could accurately and rapidly determine the urinary level of L-FABP with high-throughput, which could be useful as a new tool to predict complications in patients with kidney disease. The clinical trial was approved by Shuyang Hospital of Traditional Chinese Medicine Ethics Committee: 20,210,202-001 at February 2, 2021.
Keywords: Chemiluminescence immunoassay; Immunomagnetic bead separation; Liver fatty-acid-binding protein; Urinary biomarker.
© 2023. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.