Multicentre experience with Optimus balloon-expandable cobalt-chromium stents in congenital heart disease interventions

Open Heart. 2023 Jan;10(1):e002157. doi: 10.1136/openhrt-2022-002157.

Abstract

Objectives: To evaluate bare-metal Optimus and polytetrafluoroethylene (PTFE)-covered Optimus-CVS balloon-expandable, cobalt-chromium, hybrid cell-designed stents in congenital heart disease (CHD) interventions.

Methods: Retrospective multicentre review of patients with CHD receiving Optimus stents. Stent mechanical behaviour, clinical indications and outcomes were assessed.

Results: 183 stents (49.2% XXL/15-ZIG, 33.3% XL/12-ZIG, 17.5% L/9-ZIG) were implanted (98.9% success rate, 2.3% serious procedural complication rate) in 170 patients (57.6% men, 64.1% adults), median age 23.6 years (IQR, 15.2-39.2) and weight 63.5 kg (IQR, 47-75.7). Indications were right ventricular outflow tract stand-alone stenting or before revalvulation (62.4%), aortic coarctation treatment (15.3%), Fontan-circuit fenestration closure (12.4%) and miscellaneous (10%). 86/170 (50.6%) patients had PTFE-covered stenting (50% prophylactic). In 86/170 (50.6%) patients with stenotic lesions, median percentage of achieved stent expansion was 93.4% (IQR, 85.5%-97.7%), median gradient decreased from 28 mm Hg (IQR, 19-41) to 5 mm Hg (IQR, 1-9) (p<0.001), median vessel diameters increased from 13 mm (IQR, 7.9-17) to 18.9 mm (IQR, 15.2-22) (p<0.001) and percentage of vessel expansion was 45.2% (IQR, 19.8%-91.3%). In 30/36 (83.3%) patients with graft, median dilation of 2 mm (IQR, 2-5) above nominal diameter was achieved. Median stent shortening was 10.9% (IQR, 6.1-15.1) and was associated only with expansion diameter (OR: 0.66, 95% CI: 0.38 to 0.93). No clinically relevant fracture, stent embolisation or dysfunction occurred on a median follow-up of 9 (IQR, 4-14) months.

Conclusions: Optimus stents are effective tools for transcatheter treatment of simple and complex CHD. Optimus stents' reliable mechanical behaviour and particular covering design can promote widespread use.

Keywords: cardiac catheterisation; congenital heart disease; endovascular procedures; heart defects, congenital; outcome assessment, health care.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Aortic Coarctation*
  • Chromium
  • Cobalt
  • Female
  • Heart Defects, Congenital* / diagnosis
  • Heart Defects, Congenital* / surgery
  • Humans
  • Male
  • Polytetrafluoroethylene
  • Stents
  • Young Adult

Substances

  • Chromium
  • Cobalt
  • Polytetrafluoroethylene