Validity study of a multiscaled fusion network using single-lead electrocardiogram signals for obstructive sleep apnea diagnosis

J Clin Sleep Med. 2023 Jun 1;19(6):1017-1025. doi: 10.5664/jcsm.10466.

Abstract

Study objectives: We evaluated the validity of a squeeze-and-excitation and multiscaled fusion network (SE-MSCNN) using single-lead electrocardiogram (ECG) signals for obstructive sleep apnea detection and classification.

Methods: Overnight polysomnographic data from 436 participants at the Sleep Center of the First Affiliated Hospital of Sun Yat-sen University were used to generate a new FAH-ECG dataset comprising 260, 88, and 88 single-lead ECG signal recordings for training, validation, and testing, respectively. The SE-MSCNN was employed for detection of apnea-hypopnea events from the acquired ECG segments. Sensitivity, specificity, accuracy, and F1 scores were assigned to assess algorithm performance. We also used the SE-MSCNN to estimate the apnea-hypopnea index, classify obstructive sleep apnea severity, and compare the agreement between 2 sleep technicians.

Results: The SE-MSCNN's accuracy, sensitivity, specificity, and F1 score on the FAH-ECG dataset were 86.6%, 83.3%, 89.1%, and 0.843, respectively. Although slightly inferior to previously reported results using public datasets, it is superior to state-of-the-art open-source models. Furthermore, the SE-MSCNN had good agreement with manual scoring, such that the Spearman's correlations for the apnea-hypopnea index between the SE-MSCNN and 2 technicians were 0.93 and 0.94, respectively. Cohen's kappa scores in classifying the SE-MSCNN and the 2 sleep technicians were 0.72 and 0.78, respectively.

Conclusions: In this study, we validated the use of the SE-MSCNN in a clinical environment, and despite some limitations the network appeared to meet the performance standards for generalizability. Therefore, updating algorithms based on single-lead ECG signals can facilitate the development of novel wearable devices for efficient obstructive sleep apnea screening.

Citation: Yue H, Li P, Li Y, et al. Validity study of a multiscaled fusion network using single-lead electrocardiogram signals for obstructive sleep apnea diagnosis. J Clin Sleep Med. 2023;19(6):1017-1025.

Keywords: ECG signals; automatic scoring; deep neural networks; obstructive sleep apnea; wearable device.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrocardiography / methods
  • Humans
  • Polysomnography / methods
  • Sleep
  • Sleep Apnea Syndromes* / diagnosis
  • Sleep Apnea, Obstructive* / diagnosis