Introduction: It is not known whether the optimal atrioventricular (AVopt ) delay varies between left ventricular (LV) pacing site during endocardial biventricular pacing (BiVP) and may therefore needs consideration.
Methods: We assessed the hemodynamic AVopt in patients with chronic heart failure undergoing endocardial LV lead implantation. AVopt was assessed during atrio-BiVP with a "roving LV lead." Up to four locations were studied: mid-lateral wall, mid-septum (or a close alternative), site of greatest hemodynamic improvement, and LV lead implant site. The AVopt was compared to a fixed AV delay of 180 ms.
Results: Seventeen patients were included (12 male, aged 66.5 ± 12.8 years, ejection fraction 26 ± 7%, 16 left bundle branch block or high percentage of right ventricular pacing [RVP], QRS duration 167 ± 27 ms). In most locations (62/63), AVopt increased systolic blood pressure during BiVP compared with RVP (relative improvement 6 mmHg, interquartile range [IQR] 4-9 mmHg). Compared to a fixed AV delay, the hemodynamic improvement at AVopt was higher (1 mmHg, IQR 0.2-2.6 mmHg, p < .001). Within most patients (16/17), we observed a difference in AVopt between pacing sites (median paced AVopt 209 ms, IQR 117-250). Within this range, the hemodynamic impact of these differences was small (median loss 0.6 mmHg, IQR 0.1-2.6 mmHg).
Conclusion: Within a patient, different endocardial LV lead locations have slightly different hemodynamic AVopt which are superior to a fixed AV delay. The hemodynamic consequence of applying an optimum from a different lead location is small.
Keywords: atrioventricuar delay; cardiac resynchronization therapy; endocardial left ventricular lead; heart failure; hemodynamic optimization.