Twisting of HeartMate II outflow graft 2.5 years after implantation-HM2 is still ongoing

J Artif Organs. 2024 Mar;27(1):69-72. doi: 10.1007/s10047-023-01387-y. Epub 2023 Feb 20.

Abstract

A 47-year-old woman with dilated cardiomyopathy underwent HeartMate II (HM2) implantation as a bridge-to-transplantation. Her postoperative course was good. However, 2.5 years after surgery, the outflow graft was found to be twisted and the graft and pump was exchanged. While HeartMate 3(HM3) twisting of the outflow graft is well documented, such malfunctions in HM2 are almost unknown. Although HM2 has since been discontinued, there are a significant number of patients using HM2 who are awaiting heart transplants or destination therapy. We caution that, even with HM2, the possibility of late-phase twisting requires vigilance.

Keywords: HeartMate II; Left ventricular assist device; Outflow graft twist.

Publication types

  • Case Reports

MeSH terms

  • Female
  • Heart Failure* / surgery
  • Heart Transplantation*
  • Heart-Assist Devices*
  • Humans
  • Middle Aged
  • Retrospective Studies