Acute and long-term outcomes of quadripolar IS-4 versus bipolar IS-1 left ventricular leads in cardiac resynchronization therapy: A retrospective registry study

Pacing Clin Electrophysiol. 2023 May;46(5):365-375. doi: 10.1111/pace.14686. Epub 2023 Apr 5.

Abstract

Background: The implantation procedure of left ventricular (LV) leads and the management of cardiac resynchronization therapy (CRT) patients can be challenging. The IS-4 standard for CRT offers additional pacing vectors compared to bipolar leads (IS-1). IS-4 leads improve procedural outcome and may also result in lower adverse events during follow-up (FU) and improve clinical outcome in CRT patients. Further long-term FU data comparing the two lead designs are necessary.

Methods: In this retrospective, single-center study we included adult patients implanted with a CRT-Defibrillator (CRT-D) or CRT-Pacemaker (CRT-P) with a quadripolar (IS-4 group) or bipolar (IS-1 group) LV lead and with available ≥3 years clinical FU. The combined primary endpoint was a combination of predefined, lead-related adverse events. Secondary endpoints were all single components of the primary endpoint.

Results: Overall, 133 patients (IS-4 n = 66; IS-1 n = 67) with a mean FU of 4.03 ± 1.93 years were included. Lead-related adverse events were less frequent in patients with an IS-4 lead than with an IS-1 lead (n = 8, 12.1% vs. n = 23, 34.3%; p = .002). The secondary outcomes showed a lower rate of LV lead deactivation/explantation and LV lead dislodgement/dysfunction (4.5% vs. 22.4%; p = .003; 4.5% vs. 17.9%; p = .015, respectively) in the IS-4 patient group. Less patients suffered from unresolved phrenic nerve stimulation with an IS-4 lead (3.0% vs. 13.4%; p = .029). LV lead-related re-interventions were fewer in case of an IS-4 lead (6.1% vs. 17.9%; p = .036).

Conclusion: In this retrospective analysis, the IS-4 LV lead is associated with lower lead-related complication rates than the IS-1 lead at long-term FU.

Keywords: CRT; IS-1; IS-4.

MeSH terms

  • Adult
  • Bipolar Disorder* / therapy
  • Cardiac Resynchronization Therapy Devices
  • Cardiac Resynchronization Therapy* / methods
  • Electrodes, Implanted
  • Heart Failure*
  • Humans
  • Registries
  • Retrospective Studies
  • Treatment Outcome
  • Ventricular Dysfunction, Left* / therapy