Advanced Hybrid Closed Loop in Adult Population With Type 1 Diabetes: A Substudy From the ADAPT Randomized Controlled Trial in Users of Real-Time Continuous Glucose Monitoring

J Diabetes Sci Technol. 2024 Sep;18(5):1132-1138. doi: 10.1177/19322968231161320. Epub 2023 Mar 22.

Abstract

Background: This analysis reports the findings from a predefined exploratory cohort (cohort B) from the ADAPT (ADvanced Hybrid Closed Loop Study in Adult Population with Type 1 Diabetes) study. Adults with type 1 diabetes (T1D) with suboptimal glucose control were randomly allocated to an advanced hybrid closed-loop (AHCL) system or multiple daily injections of insulin (MDI) plus real-time continuous glucose monitoring (RT-CGM).

Methods: In this prospective, multicenter, exploratory, open-label, randomized controlled trial, 13 participants using MDI + RT-CGM and with HbA1c ≥8.0% were randomized to switch to AHCL (n = 8) or continue with MDI + RT-CGM (n = 5) for six months. Prespecified endpoints included the between-group difference in mean change from baseline in HbA1c, CGM-derived measures of glycemic control, and safety.

Results: The mean HbA1c level decreased by 1.70 percentage points in the AHCL group versus a 0.60 percentage point decrease in the MDI + RT-CGM group, with a model-based treatment effect of -1.08 percentage points (95% confidence interval [CI] = -2.17 to 0.00 percentage points; P = .0508) in favor of AHCL. The percentage of time spent with sensor glucose levels between 70 and 180 mg/dL in the study phase was 73.6% in the AHCL group and 46.4% in the MDI + RT-CGM group; model-based between-group difference of 28.8 percentage points (95% CI = 12.3 to 45.3 percentage points; P = .0035). No diabetic ketoacidosis or severe hypoglycemia occurred in either group.

Conclusions: In people with T1D with HbA1c ≥8.0%, the use of AHCL resulted in improved glycemic control relative to MDI + RT-CGM. The scale of improvement suggests that AHCL should be considered as an option for people not achieving good glycemic control on MDI + RT-CGM.

Keywords: HbA1c; automated insulin delivery; closed-loop system; diabetes; hyperglycemia; hypoglycemia; time in range.

Publication types

  • Randomized Controlled Trial
  • Multicenter Study

MeSH terms

  • Adult
  • Blood Glucose Self-Monitoring* / methods
  • Blood Glucose* / analysis
  • Blood Glucose* / drug effects
  • Continuous Glucose Monitoring
  • Diabetes Mellitus, Type 1* / blood
  • Diabetes Mellitus, Type 1* / drug therapy
  • Female
  • Glycated Hemoglobin* / analysis
  • Glycemic Control / methods
  • Humans
  • Hypoglycemic Agents* / administration & dosage
  • Hypoglycemic Agents* / adverse effects
  • Insulin Infusion Systems*
  • Insulin* / administration & dosage
  • Male
  • Middle Aged
  • Prospective Studies

Substances

  • Blood Glucose
  • Insulin
  • Hypoglycemic Agents
  • Glycated Hemoglobin
  • hemoglobin A1c protein, human