Purpose: To study how the quantitative parameters of 18F-FDG PET imaging change with the emission scan duration (ESD) and the body-mass-index (BMI) in phantom and patients on a time-of-flight (TOF)-PET/CT system.
Methods: The image-quality phantom with (b-NEMA-IQ, BMI = 29.2 kg/m2) and without (NEMA-IEC, BMI = 21.4 kg/m2) a 'belt' of water-bags was filled with 18F-FDG activities to obtain nominal standardized uptake values (SUV) of 19, 8 and 5. Patients with BMI ≤ 25 kg/m2 (L-BMI) and BMI > 25 kg/m2 (H-BMI) were enrolled in this study. Phantom and patients underwent list-mode PET acquisition at 120 s/bed-position. Images reconstructed with clinical protocol and different ESD (120, 90, 75, 60, 45, 30 s) were analysed for comparison of maximum SUV (SUVmax), maximum standardized uptake value lean-body-mass corrected (SULmax) and noise.
Results: 79 oncologic patients (45 L-BMI, 44 H-BMI) were analysed. From 90 s to 30 s, an increasing variation of SUVmax and SULmax with respect to the reference 120 s time was observed, from 18% to 60% and from 16% to 37% for phantom and patients, respectively. SUVmax values were significantly higher (+50%) in b-NEMA-IQ than NEMA-IQ phantom and in H-BMI (+33%) than L-BMI patients. No significant difference was found in SULmax for the two BMI categories in both phantom and patients. CV values decreased when increasing ESD, being higher in H-BMI patients (0.13-0.25) and b-NEMA-IQ phantom (0.15-0.28) than in L-BMI patients (0.11-0.21) and NEMA-IQ phantom (0.11-0.20).
Conclusions: Reduction of ESD may severely impact on the variations of SUVmax and SULmax in 18F-FDG PET/CT imaging. This study confirms recommendations of using SUL for lesion uptake quantification, being unaffected by BMI variation.
Keywords: Image quality; PET/CT; SUL; SUV.