The stopping-power ratio (SPR) of body tissues relative to water depends on the particle energy and mean excitation energy (I value) of the tissues. Effective energies to minimize the range error in proton therapy and ion beam therapy with helium, carbon, oxygen, and neon ions and elemental I values have been updated in recent studies. We investigated the effects of these updates on SPR estimation for computed tomography-based treatment planning. The updates led to an increase of up to 0.5% in the SPRs of soft tissues, whereas they led to a decrease of up to 1.9% in the SPRs of bone tissues compared with the current clinical settings. For 44 proton beams planned for 15 randomly sampled patients, the mean water-equivalent target depth change was - 0.2 mm with a standard deviation of 0.2 mm. The maximum change was - 0.6 mm, which we consider to be insignificant in clinical practice.
Keywords: Charged-particle therapy; Range estimation; Stopping-power ratio; Treatment planning.