The Woven EndoBridge (WEB) (MicroVention/Terumo) device is a treatment option for wideneck bifurcation aneurysms. An uncommon adverse effect is WEB device migration. While certain bailout strategies for WEB recovery have been described, there is still a paucity of information on optimal strategies to maximize both short and long-term post-operative outcomes. We add 2 cases at our institution to the existing literature of WEBectomy in the setting of complicated intracranial aneurysm treatment. We discuss the long-term imaging outcomes with additional fluoroscopy video demonstrating our technique. Our findings reflect a clear benefit for the use of the Amplatz GooseneckTM microsnare (Medtronic) device as a means of WEB recovery, coupled with potential stent-assisted WEB embolization to remove the aneurysm from the parent circulation, while minimizing recurrence and thromboembolic complications.
Keywords: Cerebral aneurysm; Microsnare; WEB; Woven endobridge.