Factors affecting electrogram sensing in an insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis

Heart Rhythm. 2023 Sep;20(9):1297-1306. doi: 10.1016/j.hrthm.2023.05.034. Epub 2023 Jun 13.

Abstract

Background: Fidelity of electrogram sensing may reduce false alerts from an insertable cardiac monitor (ICM).

Objective: The purpose of this study was to assess the effect of vector length, implant angle, and patient factors on electrogram sensing using surface electrocardiogram (ECG) mapping.

Methods: Twelve separate precordial single-lead surface ECGs were acquired from 150 participants at 2 interelectrode distances (75 and 45 mm), at 3 vector angles (vertical, oblique, and horizontal), and in 2 postures (upright and supine). A subset of 50 patients also received a clinically indicated ICM implant in 1:1 ratio (Reveal LINQ [Medtronic, Minneapolis, MN]/BIOMONITOR III [Biotronik, Berlin, Germany]). All ECGs and ICM electrograms were analyzed by blinded investigators using DigitizeIt software (V2.3.3, Braunschweig, Germany). The P-wave visibility threshold was set at > 0.015 mV. Logistic regression was used to identify factors affecting P-wave amplitude.

Results: A total of 1800 tracings from 150 participants (44.5% [n = 68] female; median age 59 [35-73] years) were assessed. The median P- and R-wave amplitudes were 45% and 53% larger with vector lengths of 75 and 45 mm, respectively (P < .001 for both). The oblique orientation yielded the best P- and R-wave amplitudes, while posture change did not affect P-wave amplitude. Mixed effects modeling found that visible P-waves occur more frequently with a vector length of 75 mm than with 45 mm (86% vs 75%, respectively; P < .0001). A longer vector length improved both P-wave amplitude and visibility in all body mass index categories. There was a moderate correlation of P- and R-wave amplitudes from the ICM electrograms to those from surface ECG recordings (intraclass correlation coefficient 0.74 and 0.80, respectively).

Conclusion: Longer vector length and oblique implant angle yielded the best electrogram sensing and are relevant considerations for ICM implantation procedures.

Keywords: ECG mapping; Electrogram sensing; Insertable cardiac monitor; P-wave visibility; Vector length.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrocardiography* / methods
  • Electrocardiography, Ambulatory* / methods
  • Female
  • Germany
  • Humans
  • Middle Aged
  • Prostheses and Implants
  • Software

Associated data

  • ANZCTR/ACTRN12620000140954