ACT5 Electrical Impedance Tomography System

IEEE Trans Biomed Eng. 2024 Jan;71(1):227-236. doi: 10.1109/TBME.2023.3295771. Epub 2023 Dec 22.

Abstract

Objective: This article introduces the Adaptive Current Tomograph 5 (ACT5) Electrical Impedance Tomography (EIT) system. ACT5 is a 32 electrode applied-current multiple-source EIT system that can display real-time images of conductivity and susceptivity at 27 frames per second. The adaptive current sources in ACT5 can apply fully programmable current patterns with frequencies varying from 5 kHz to 500 kHz. The system also displays real-time ECG readings during the EIT imaging process.

Methods: The hardware and software design and specifications are presented, including the current source design, FPGA hardware, safety features, calibration, and shunt impedance measurement.

Results: Images of conductivity and susceptivity are presented from ACT5 data collected on tank phantoms and a human subject illustrating the system's ability to provide real-time images of pulsatile perfusion and ECG traces.

Significance: The portability, high signal-to-noise ratio, and flexibility of applied currents over a wide range of frequencies enable this instrument to be used to obtain useful human subject data with relative clinical ease.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Computers
  • Electric Conductivity
  • Electric Impedance
  • Humans
  • Tomography* / methods
  • Tomography, X-Ray Computed*