The Effectiveness of Ultraviolet Smart D60 in Reducing Contamination of Flexible Fiberoptic Laryngoscopes

Laryngoscope. 2023 Dec;133(12):3512-3519. doi: 10.1002/lary.30869. Epub 2023 Jul 24.

Abstract

Objective: To compare the effectiveness of disinfection protocols utilizing a ultraviolet (UV) Smart D60 light system with Impelux™ technology with a standard Cidex ortho-phthalaldehyde (OPA) disinfection protocol for cleaning flexible fiberoptic laryngoscopes (FFLs).

Methods: Two hundred FFLs were tested for bacterial contamination after routine use, and another 200 FFLs were tested after disinfection with one of four methods: enzymatic detergent plus Cidex OPA (standard), enzymatic detergent plus UV Smart D60, microfiber cloth plus UV Smart D60, and nonsterile wipe plus UV Smart D60. Pre- and post-disinfection microbial burden levels and positive culture rates were compared using Kruskal-Wallis ANOVA and Fisher's two-sided exact, respectively.

Results: After routine use, approximately 56% (112/200) of FFLs were contaminated, with an average contamination level of 9,973.7 ± 70,136.3 CFU/mL. The standard reprocessing method showed no positive cultures. The enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV methods yielded contamination rates of 4% (2/50), 6% (3/50), and 12% (6/50), respectively, with no significant differences among the treatment groups (p > 0.05). The pre-disinfection microbial burden levels decreased significantly after each disinfection technique (p < 0.001). The average microbial burden recovered after enzymatic plus UV, microfiber plus UV, and nonsterile wipe plus UV were 0.40 CFU/mL ± 2, 0.60 CFU/mL ± 2.4, and 12.2 CFU/mL ± 69.5, respectively, with no significant difference among the treatment groups (p > 0.05). Micrococcus species (53.8%) were most frequently isolated, and no high-concern organisms were recovered.

Conclusion: Disinfection protocols utilizing UV Smart D60 were as effective as the standard chemical disinfection protocol using Cidex OPA.

Level of evidence: NA Laryngoscope, 133:3512-3519, 2023.

Keywords: UV disinfection; flexible fiberoptic laryngoscope; semi-critical device.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Detergents
  • Disinfection / methods
  • Equipment Contamination / prevention & control
  • Glutaral
  • Humans
  • Laryngoscopes* / microbiology
  • o-Phthalaldehyde

Substances

  • Glutaral
  • Detergents
  • o-Phthalaldehyde

Grants and funding