Our institutional approach to performing periodic constancy checks on dose calculations for the Varian Eclipse treatment planning system is described, including details of how the Radformation ClearCheck software is leveraged for highly efficient results analysis. Photon- and electron-specific synthetic phantoms are created which allow all modality-specific fields to be spatially separated and calculated in a single plan, with field-specific structures encompassing individual irradiated volumes facilitating field-specific dose distribution evaluations. Plan comparison templates in ClearCheck allow for the near-instantaneous evaluation of 116 and 165 individual calculation metrics for photon and electron fields, respectively. The methodology is further applied to the Radformation ClearCalc finite-size pencil beam and electron algorithms for secondary algorithm validation. The total time required for constancy evaluations of 6 primary algorithms and 4 secondary algorithms is less than one hour, with much of that time used for passive dose calculation.
Keywords: Quality assurance; Treatment planning systems.
Copyright © 2023 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.