Silicone urinary catheters infused with silicone liquid offer an effective alternative to antibiotic coatings, reducing microbial adhesion while decreasing bladder colonization and systemic dissemination. However, loss of free silicone liquid from the surface into the host system is undesirable. To reduce the potential for liquid loss, free silicone liquid was removed from the surface of liquid-infused catheters by either removing excess liquid from fully infused samples or by partial infusion. The effect on bacterial and host protein adhesion was then assessed. Removing the free liquid from fully infused samples resulted in a ~64% decrease in liquid loss into the environment compared to controls, with no significant increase in deposition of the host protein fibrinogen or the adhesion of the common uropathogen Enterococcus faecalis. Partially infusing samples decreased liquid loss as total liquid content decreased, with samples infused to 70-80% of their maximum capacity showing a ~85% reduction in liquid loss compared to fully infused controls. Furthermore, samples above 70% infusion showed no significant increase in fibrinogen or E. faecalis adhesion. Together, the results suggest that eliminating free liquid layer, mechanically or through partial infusion, can reduce liquid loss from liquid-infused catheters while preserving functionality.
Keywords: anti-fouling; catheter-associated urinary tract infections; catheters; liquid infusion; medical devices; silicone liquid.