High-resolution Diffusion-weighted Imaging of the Prostate Using Multiplexed Sensitivity-encoding: Comparison with the Conventional and Reduced Field-of-view Techniques

Magn Reson Med Sci. 2025 Jan 1;24(1):58-65. doi: 10.2463/mrms.mp.2023-0039. Epub 2023 Oct 28.

Abstract

Purpose: To compare objective and subjective image quality, lesion conspicuity, and apparent diffusion coefficient (ADC) of high-resolution multiplexed sensitivity-encoding diffusion-weighted imaging (MUSE-DWI) with conventional DWI (c-DWI) and reduced FOV DWI (rFOV-DWI) in prostate MRI.

Methods: Forty-seven patients who underwent prostate MRI, including c-DWI, rFOV-DWI, and MUSE-DWI, were retrospectively evaluated. SNR and ADC of normal prostate tissue and contrast-to-noise ratio (CNR) and ADC of prostate cancer (PCa) were measured and compared between the three sequences. Image quality and lesion conspicuity were independently graded by two radiologists using a 5-point scale and compared between the three sequences.

Results: The SNR of normal prostate tissue was significantly higher with rFOV-DWI than with the other two DWI techniques (P ≤ 0.01). The CNR of the PCa was significantly higher with rFOV-DWI than with MUSE-DWI (P < 0.05). The ADC of normal prostate tissue measured by rFOV-DWI was lower than that measured by MUSE-DWI and c-DWI (P < 0.01), while there was no difference in the ADC of cancers. In the qualitative analysis, MUSE-DWI showed significantly higher scores than rFOV-DWI and c-DWI for visibility of anatomy and overall image quality in both readers, and significantly higher scores for distortion in one of the two readers (P < 0.001). There was no difference in lesion conspicuity between the three sequences.

Conclusion: High-resolution MUSE-DWI showed higher image quality and reduced distortion compared to c-DWI, while maintaining a wide FOV and similar ADC quantification, although no difference in lesion conspicuity was observed.

Keywords: apparent diffusion coefficient; diffusion-weighted imaging; magnetic resonance imaging; prostate cancer.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Diffusion Magnetic Resonance Imaging* / methods
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods
  • Image Processing, Computer-Assisted / methods
  • Male
  • Middle Aged
  • Prostate* / diagnostic imaging
  • Prostate* / pathology
  • Prostatic Neoplasms* / diagnostic imaging
  • Prostatic Neoplasms* / pathology
  • Reproducibility of Results
  • Retrospective Studies
  • Sensitivity and Specificity
  • Signal-To-Noise Ratio