Noise equivalent count density (NEC density ) is often used to evaluate the image quality of whole-body fluorodeoxyglucose tomography tests. However, this index is calculated using the patient volume, which is difficult to obtain at every facility. In this study, we proposed new image quality indices that can be evaluated at all facilities. In total, 94 patients were enrolled in the study. The correlations of patients' body weight and BMI with volume were examined. New image quality indices normalized by body weight and BMI were defined as NEC bw and NEC bmi , respectively. Correlations between NEC bw , NEC bmi , and NEC density were examined. Further, the correlations between these two new indices and visual scores were evaluated. Good correlations were observed between volume and body weight (r = 0.861, P < 0.001) and between volume and BMI (r = 0.728, P < 0.001). NEC bw and NEC bmi correlated well with NEC density (r = 0.954 for NEC bw and r = 0.897 for NEC bmi , P < 0.001). These correlations improved when the examined bed positions were set to the same number. Additionally, the correlations of visual scores with NEC bw and NEC bmi were similar to those between the visual score and NEC density . Our investigation indicated that the newly proposed image quality metrics, NEC bw and NEC bmi , were easily calculated and as useful as NEC density for evaluating image quality when subjects had similar physiques.
Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.