Background: Robotic assistance in total hip arthroplasty (RA-THA) has been shown to minimize laterality-based differences in acetabular cup positioning.
Objective: To determine if the use of a novel, fluoroscopy-based RA-THA system mitigates differences in acetabular cup placement between left (L) and right (R) side hip procedures, when compared to manual, fluoroscopic-assisted technique.
Methods: We conducted a retrospective review of 106 consecutive mTHA (40 L/66 R) and 102 RA-THA (48 L/54 R) primary direct anterior approach procedures. All cases were performed by a single right-hand-dominant surgeon, for a pre-operative diagnosis of osteoarthritis, avascular necrosis, or rheumatoid arthritis. Outcomes included acetabular cup inclination and anteversion, and the proportion of cups within the Lewinnek safe-zone.
Results: The average inclination of mTHA L cases was smaller than that of mTHA R cases (41.10∘± 7.38 vs. 43.97∘± 6.27; p= 0.04). For RA-THA, L and R cup angles were similar. There were fewer overall mTHA hips within the Lewinnek safe-zone compared to RA-THA (0.59 vs. 0.78; p= 0.003), as well as fewer mTHA R cases than RA-THA R cases (0.59 vs. 0.80; p= 0.03) within safe zone.
Conclusion: Use of a novel, fluoroscopy-based robotic system mitigates laterality-based differences in acetabular cup placement that were observed in a manual, fluoroscopic-assisted cohort.