Influence of ionization volume and sample gas flow rate on separation power in gas chromatography-ion mobility spectrometry

J Chromatogr A. 2024 Jan 4:1713:464506. doi: 10.1016/j.chroma.2023.464506. Epub 2023 Nov 13.

Abstract

In this work, the influence of the sample gas flow rate and the ionization region volume of an ion mobility spectrometer (IMS) used as a detector in gas chromatography (GC) on GC-IMS peak shape has been investigated. Therefore, a drift tube IMS with a field-switching ion shutter, a defined ionization region volume and an ultra-violet radiation source was used. To identify the influence of the sample gas flow rate entering the ionization region (equals the GC carrier gas flow rate if no further make-up gas is used) and the ionization region volume on peak broadening and signal intensity, different sample volumes as they would elute from a GC were tested at a variety of sample gas flow rates at a given ionization region volume. The results clearly show that for low sample gas flow rates a depletion of sample molecules in the ionization region leads to a significant decrease in effective detector volume but also to reduced signal intensities. Therefore, for optimal performance of a GC-IMS, the optimal operating point of the GC should match the flow range, where the IMS provides the best compromise between signal-to-noise ratio and peak broadening.

Keywords: Effective detector volume; GC-IMS; Ionization volume; Peak broadening.

MeSH terms

  • Gas Chromatography-Mass Spectrometry / methods
  • Ion Mobility Spectrometry* / methods
  • Signal-To-Noise Ratio