We report a case of PreserFlo MicroShunt (PFM) dislocation following a postsurgical needling procedure. A 58-year-old woman underwent PFM implantation for exfoliation glaucoma in her left eye (OS). There were no intraoperative complications. Preoperatively, her best-corrected visual acuity (BCVA) was 0.6, and her intraocular pressure (IOP) was 25 mmHg with three antiglaucoma medications in the OS. On postoperative day 21, the IOP was 21 mmHg OS, and the filtration bleb had shrunk. A needling procedure was performed using a sharp 26-gauge needle to lower the IOP. On postoperative day 29, the BCVA was 0.02, and the IOP was 60 mmHg OS. Gonioscopy revealed no device tip in the anterior chamber, and peripheral anterior synechia was observed at the site of PFM insertion. Anterior segment optical coherence tomography showed a dislocated device in the subconjunctival space. On postoperative day 35, the dislocated PFM was removed, and a new device was inserted. Following the reoperation, no further complications were observed, and bleb formation was obtained. In conclusion, like other glaucoma filtering surgeries, PFM may require postsurgical needling procedures. Needling procedures may cause PFM dislocation and IOP rise, resulting in the requirement for further IOP-reducing procedures.
Keywords: bleb needling; device dislocation; glaucoma surgery; preserflo microshunt; surgical complication.
Copyright © 2023, Murakami et al.