SARS-CoV-2, influenza A/B virus (IAV/IBV), and respiratory syncytial virus (RSV) are among the common viruses causing acute respiratory infections. Clinical diagnosis to differentiate these viruses is challenging due to similar clinical presentations; thus, laboratory-based real-time RT PCR is the gold standard for diagnosis. This retrospective study aimed to evaluate the diagnostic performance of STANDARD M10 Flu/RSV/SARS-CoV-2 (SD Biosensor Inc., Seoul, Korea) using archived positive and negative respiratory samples for SARS-CoV-2, IAV, IBV, and RSV. A total of 322 respiratory samples were tested, comprising 215 positive samples (49 SARS-CoV-2, 48 IAV, 53 IBV, 65 RSV) and 107 negative samples. All samples were tested with both STANDARD M10 and compared to either Xpert Xpress SARS-CoV-2 or Xpert Xpress Flu/RSV (Cepheid, Sunnyvale, CA, USA). The sensitivity, specificity, positive predictive value, and negative predictive value rates of STANDARD M10 were very similar to Xpert Xpress SARS-CoV-2 or Xpert Xpress Flu/RSV ranges for each virus (98-100%). The duration of testing and workflows were similar. The overall agreement was 99.4%, including 99.1% agreement for positive samples and 100% agreement for negative samples. In conclusion, the STANDARD M10 point-of-care test is suitable for rapid simultaneous detection of SARS-CoV-2, IAV, IBV, and RSV.
Keywords: Malaysia; SARS-CoV-2; diagnostic accuracy; influenza A virus; influenza B virus; point-of-care testing; rapid RT-PCR; respiratory syncytial virus.