In this paper, we report on a fluorescent and colorimetric system for measuring the dilution of capillary blood released by a needle-free jet injector. Jet injection uses a high-speed liquid jet to penetrate tissue, and in the process can release capillary blood that can be collected for performing blood tests. In this way, blood sampling can be performed without the use of a lancet. However, any injectate that mixes with the collected blood dilutes the sample and may significantly impact subsequent analyses. By adding the fluorescent marker indocyanine green to the injected liquid, the fraction of injectate mixed into the collected blood can be measured. The incorporation of colorimetry allows our system to also correct for the impact of hematocrit on fluorescence. The results from this system show that it can determine the dilution of blood that has been diluted by up to 10 %, the upper limit of dilution typically observed in lancet-free blood sampling via jet injection.Clinical Relevance- Blood samples can be collected by jet injection without significant dilution, avoiding the need for lancing.