Purpose: Mechanical circulatory support (MCS) using a venoarterial extracorporeal membrane oxygenation (VA-ECMO) device or a catheter-type heart pump (Impella) is critical for the rescue of patients with severe cardiogenic shock. However, these MCS devices require large-bore cannula access (14-Fr and larger) at the femoral artery or vein, which often requires surgical decannulation.
Methods: In this retrospective study, we evaluated post-closure method using a percutaneous suture-mediated vascular closure system, Perclose ProGlide/ProStyle (Abbott Vascular, Lake Bluff, IL, Perclose), as an alternative procedure for MCS decannulation. Closure of 83 Impella access sites and 68 VA-ECMO access sites performed using Perclose or surgical method between January 2018 and March 2023 were evaluated.
Results: MCS decannulation using Perclose was successfully completed in all access sites without surgical hemostasis. The procedure time of ProGlide was shorter than surgical decannulation for both Impella and VA-ECMO (13 min vs. 50 min; p < 0.001, 21 min vs. 65 min; p < 0.001, respectively). There were no significant differences in the 30-day survival rate and major adverse events by decannulation including arterial dissection requiring endovascular treatment, hemorrhage requiring a large amount of red blood cell transfusion, and access site infection.
Conclusion: Our results suggest that the post-closure technique using the percutaneous suture-mediated closure system appears to be a safe and effective method for large-bore MCS decannulation.
Keywords: Impella; Large-bore decannulation; Mechanical circulatory support; Percutaneous suture-mediated vascular closure device.
Copyright © 2023 The Authors. Published by Elsevier Inc. All rights reserved.