Native myocardial T1 mapping: influence of spatial resolution on quantitative results and reproducibility

Quant Imaging Med Surg. 2024 Jan 3;14(1):20-30. doi: 10.21037/qims-23-943. Epub 2023 Nov 14.

Abstract

Background: Myocardial mapping techniques can be used to quantitatively assess alterations in myocardial tissue properties. This study aims to evaluate the influence of spatial resolution on quantitative results and reproducibility of native myocardial T1 mapping in cardiac magnetic resonance imaging (MRI).

Methods: In this cross-sectional study with prospective data collection between October 2019 and February 2020, 50 healthy adults underwent two identical cardiac MRI examinations in the radiology department on the same day. T1 mapping was performed using a MOLLI 5(3)3 sequence with higher (1.4 mm × 1.4 mm) and lower (1.9 mm × 1.9 mm) in-plane spatial resolution. Global quantitative results of T1 mapping were compared between high-resolution and low-resolution acquisitions using paired t-test. Intra-class correlation coefficient (ICC) and Bland-Altman statistics (absolute and percentage differences as means ± SD) were used for assessing test-retest reproducibility.

Results: There was no significant difference between global quantitative results acquired with high vs. low-resolution T1 mapping. The reproducibility of global T1 values was good for high-resolution (ICC: 0.88) and excellent for low-resolution T1 mapping (ICC: 0.95, P=0.003). In subgroup analyses, inferior test-retest reproducibility was observed for high spatial resolution in women compared to low spatial resolution (ICC: 0.71 vs. 0.91, P=0.001) and heart rates >77 bpm (ICC: 0.53 vs. 0.88, P=0.004). Apical segments had higher T1 values and variability compared to other segments. Regional T1 values for basal (ICC: 0.81 vs. 0.89, P=0.023) and apical slices (ICC: 0.86 vs. 0.92, P=0.024) showed significantly higher reproducibility in low-resolution compared to high-resolution acquisitions but without differences for midventricular slice (ICC: 0.91 vs. 0.92, P=0.402).

Conclusions: Based on our data, we recommend a spatial resolution on the order of 1.9 mm × 1.9 mm for native myocardial T1 mapping using a MOLLI 5(3)3 sequence at 1.5 T particularly in individuals with higher heart rates and women.

Keywords: Cardiac magnetic resonance imaging; T1; mapping; reproducibility; spatial resolution.