Aims: The left atrial posterior wall is a potential ablation target in patients with recurrent atrial fibrillation despite durable pulmonary vein isolation or in patients with roof-dependent atrial tachycardia (AT). Pulsed-field ablation (PFA) offers efficient and safe posterior wall ablation (PWA), but available data are scarce.
Methods and results: Consecutive patients undergoing PWA using PFA were included. Posterior wall ablation was performed using a pentaspline PFA catheter and verified by 3D-electroanatomical mapping. Follow-up was performed using 7-day Holter ECGs 3, 6, and 12 months after ablation. Recurrence of any atrial arrhythmia lasting more than 30 s was defined as failure. Lesion durability was assessed during redo procedures. Posterior wall ablation was performed in 215 patients (70% males, median age 70 [IQR 61-75] years, 67% redo procedures) and was successful in all patients (100%) by applying a median of 36 (IQR 32-44) PFA lesions. Severe adverse events were cardiac tamponade and vascular access complication in one patient each (0.9%). Median follow-up was 7.3 (IQR 5.0-11.8) months. One-year arrhythmia-free outcome in Kaplan-Meier analysis was 53%. A redo procedure was performed in 26 patients (12%) after a median of 6.9 (IQR 2.4-11) months and showed durable PWA in 22 patients (85%) with only minor lesion regression. Among four patients with posterior wall reconnection, three (75%) presented with roof-dependent AT.
Conclusion: Posterior wall ablation with this pentaspline PFA catheter can be safely and efficiently performed with a high durability observed during redo procedures. The added value of durable PWA for the treatment of atrial fibrillation remains to be evaluated.
Keywords: Atrial fibrillation; Posterior wall ablation; Pulmonary vein isolation; Pulsed-field ablation; Reconnection.
© The Author(s) 2024. Published by Oxford University Press on behalf of the European Society of Cardiology.