Background: Ciliary beat frequency (CBF) is crucial in mucociliary clearance. High-speed video analysis (HSVA) is commonly used to measure CBF but lacks standardization. We compared visual observation and computer-assisted calculation using fast Fourier transformation (FFT) in freshly collected bronchial ciliary epithelial cells and cultured cells.
Methods: Bronchial epithelial cells were obtained from 12 patients who required bronchoscopic examination. Eighty-five videos of ciliary movement of freshly collected and cultured cells were recorded and used to calculate CBF using manual observation, region of interest (ROI) selection, and whole-field analysis.
Results: CBF measured by the ROI selection method strongly correlated with that measured using manual observation, especially in freshly collected cells. However, 27.8% of the manual observation method values were doubled in the ROI selection method, probably because a round trip of cilia was calculated as two cycles and needed to be corrected to 1/2 value. Upon increasing the number of ROIs, the results of the ROI selection method came closer to that of WFA.
Conclusions: Computer-assisted calculation using FFT can aid in measuring CBF; however, current methods require visual confirmation. Further automated evaluation techniques are needed to establish more standardized and generalized CBF measurement methods using HSVA.
Keywords: Air-liquid interface culture; Ciliary beat frequency; Fast fourier transformation; High-speed video analysis.