Comparison of Locked Plates and Blade Plates for Varus Osteotomy of the Proximal Femur by the Finite Element Method

Rev Bras Ortop (Sao Paulo). 2024 Mar 21;59(1):e107-e118. doi: 10.1055/s-0043-1775889. eCollection 2024 Feb.

Abstract

Objective: The present study compared the difference in load and pressure distribution behavior of the blade plate and locked plate for varus osteotomy of the proximal femur per the finite element method. Methods: Modeling was performed by scanning a medium-sized left femur with medial valgus deformity made of polyurethane. Results: The stiffness of the locked plate is higher compared with that of the blade plate. However, this difference was not significant. In addition, the locked plate has proximal locking screws to ensure that the bending moments on the screws are smaller during loading. Conclusion: In summary, both plates are well-established and effective. However, the study using the finite element method plays a fundamental role in understanding the load and pressure distribution of the implant. Moreover, it opens up new possibilities for further studies, including surgical proposals and customized implant materials.

Keywords: bone plates; femur; orthopedics; traumatology.